Scientific publications

Validation of the ALEX microarray platform for specific IgE detection of respiratory and plant-food allergens

Dec 1, 2021 | Magazine: Annals of Allergy, Asthma & Immunology

Paola Leonor Quan  1 , Marina Sabaté-Brescó  2 , Carmen Mariana D'Amelio  3 , Mariona Pascal  4 , Blanca Esther García  5 , Gabriel Gastaminza  3 , Natalia Blanca-López  6 , Maria Isabel Alvarado  7 , Francisco Javier Fernández  8 , Carmen Moya  9 , Joan Bartra  4 , Marta Ferrer  3 , Maria Jose Goikoetxea 3


Background: As the use of multiplex specific IgE (sIgE) detection methods becomes increasingly widespread, proper comparative validation assessments of emerging new platforms are vital.

Objective: To assess the clinical and technical performance of a newly introduced microarray platform, ALEX (MacroArray Diagnostics), in the diagnosis of pollen (cypress, grass, olive), dust mite Dermatophagoides pteronyssinus, Alternaria alternata, fruit (apple, peach) and nut (walnut, hazelnut and peanut) allergy, and to compare it to that of the ImmunoCAP ISAC 112 microarray and the Immuno CAP singleplex method (ThermoFisher Scientific).

Methods: We enrolled 153 allergic patients and 16 non-atopic controls. sIgE assays were conducted using ISAC112, ALEX version 2 (ALEX2), and ImmunoCAP for whole extracts and major components. Technical validation of ALEX2 was performed by measuring repeatability and inter-assay, inter-batch and inter-lab reproducibility.

Results: When measured globally (detection by one or more allergen components), ALEX2 showed adequate sensitivity and specificity for most of the allergens studied, comparable in general to that of ISAC112 (except for olive pollen and walnut) and similar to that of ImmunoCAP whole extract measurements.

Component-by-component analysis showed comparable results for all techniques, except for Ole e 1 and Jug r 3 in both ISAC112 and ImmunoCAP comparisons, and Alt a 1, when compared with ISAC112. Continuous sIgE levels correlate with sIgE by ImmunoCAP. Good reproducibility and repeatability were observed for ALEX2.

Conclusion: ALEX2 shows sound technical performance, and adequate diagnostic capacity, comparable in general to that of ISAC112 and ImmunoCA.

CITATION  Ann Allergy Asthma Immunol. 2021 Dec 1;S1081-1206(21)01288-6. doi: 10.1016/j.anai.2021.11.019.