Reproducibility and clinical relevance of the ocular response analyzer in nonoperated eyes: corneal biomechanical and tonometric implications

Moreno-Montañés J [ES], Maldonado MJ, García N, Mendiluce L, García-Gómez PJ, Seguí-Gómez M.
Department of Ophthalmology, Clínica Universitaria, Universidad de Navarra, Pamplona, Spain.

Revisão:Investigative Ophthalmology and Visual Sciences

Data: 1/Mar/2008

Oftalmologia [ES]

PURPOSE
To assess the reproducibility of the ocular response analyzer (ORA) in nonoperated eyes and the impact of corneal biomechanical properties on intraocular pressure (IOP) measurements in normal and glaucomatous eyes.

METHODS
In the reliability study, two independent examiners obtained repeated ORA measurements in 30 eyes. In the clinical study, the examiners analyzed ORA and IOP-Goldmann values from 220 normal and 42 glaucomatous eyes. In both studies, Goldmann-correlated IOP measurement (IOP-ORAg), corneal-compensated IOP (IOP-ORAc), corneal hysteresis (CH), and corneal resistance factor (CRF) were evaluated. IOP differences of 3 mm Hg or greater between the IOP-ORAc and IOP-ORAg were considered outcome significant.

RESULTS
Intraexaminer intraclass correlation coefficients and interexaminer concordance correlation coefficients ranged from 0.78 to 0.93 and from 0.81 to 0.93, respectively, for all parameters. CH reproducibility was highest, and the IOP-ORAg readings were lowest. The median IOP was 16 mm Hg with the Goldmann tonometer, 14.5 mm Hg with IOP-ORAg (P < 0.001), and 15.7 mm Hg with IOP-ORAc (P < 0.001). Outcome-significant results were found in 77 eyes (29.38%). The IOP-ORAc, CH, and CRF were correlated with age (r = 0.22, P = 0.001; r = -0.23, P = 0.001; r = -0.14, P = 0.02, respectively), but not the IOP-ORAg or IOP-Goldmann.

CONCLUSIONS
The ORA provides reproducible corneal biomechanical and IOP measurements in nonoperated eyes. Considering the effect of ORA, corneal biomechanical metrics produces an outcome-significant IOP adjustment in at least one quarter of glaucomatous and normal eyes undergoing noncontact tonometry. Corneal viscoelasticity (CH) and resistance (CRF) appear to decrease minimally with increasing age in healthy adults.

CITAÇÃO DO ARTIGO  Invest Ophthalmol Vis Sci. 2008 Mar;49(3):968-74

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra