Publicaciones científicas

Utility Of POC Xpert HIV-1 Tests For Detection-Quantification Of Complex HIV Recombinants Using Dried Blood Spots From Kinshasa, D. R. Congo

Rubio-Garrido M (1), Ndarabu A (2), Reina G (3), Barquín D (4), Fernández-Alonso M (4), Carlos S (5), Holguín Á (1)

(1) HIV-1 Molecular Epidemiology Laboratory, Microbiology and Parasitology Department, University Hospital Ramón y Cajal-IRYCIS and CIBEREsp-RITIP, Madrid, 28034, Spain.
(2) Monkole Hospital, Kinshasa, Democratic Republic of the Congo.
(3) Microbiology Department, Clínica Universidad de Navarra, Navarra Institute for Health Research (IdiSNA), Institute of Tropical Health, Universidad de Navarra (ISTUN), Pamplona, 31008, Spain.
(4) Microbiology Department, Clínica Universidad de Navarra, Navarra Institute for Health Research (IdiSNA), Institute of Tropical Health, Universidad de Navarra (ISTUN), Pamplona, 31008, Spain.
(5) Department of Preventive Medicine and Public Health, Institute of Culture and Society (ICS), Institute of Tropical Health (ISTUN), Universidad de Navarra, Navarra Institute for Health Research (IdiSNA), Pamplona, 31008, Spain.

Revista: Sciencitific Reports

Fecha: 05-abr-2019

Microbiología Clínica

RESUMEN

Point-of-Care (POC) molecular assays improve HIV infant diagnosis and viral load (VL) quantification in resource-limited settings.

We evaluated POC performance in Kinshasa (Democratic Republic of Congo), with high diversity of HIV-1 recombinants. In 2016, 160 dried blood samples (DBS) were collected from 85 children (60 HIV-, 18 HIV+, 7 HIV-exposed) and 75 HIV+ adults (65 treated, 10 naive) at Monkole Hospital (Kinshasa). We compared viraemia with Cepheid-POC-Xpert-HIV-1VL and the non-POC-COBAS®AmpliPrep/COBAS®TaqMan®HIV-1-Testv2 in all HIV+, carrying 72.4%/7.2% HIV-1 unique/complex recombinant forms (URF/CRF).

HIV-1 infection was confirmed in 14 HIV+ children by Cepheid-POC-Xpert-HIV-1Qual and in 70 HIV+ adults by both Xpert-VL and Roche-VL, identifying 8 false HIV+ diagnosis performed in DRC (4 adults, 4 children). HIV-1 was detected in 95.2% and 97.6% of 84 HIV+ samples by Xpert-VL and Roche-VL, respectively. Most (92.9%) HIV+ children presented detectable viraemia by both VL assays and 74.3% or 72.8% of 70 HIV+ adults by Xpert or Roche, respectively. Both VL assays presented high correlation (R2 = 0.89), but showing clinical relevant ≥0.5 log VL differences in 15.4% of 78 cases with VL within quantification range by both assays.

This is the first study confirming the utility of Xpert HIV-1 tests for detection-quantification of complex recombinants currently circulating in Kinshasa.

CITA DEL ARTÍCULO  Sci Rep. 2019 Apr 5;9(1):5679. doi: 10.1038/s41598-019-41963-y

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra