Evaluation of RETICs Glaucoma Diagnostic Calculators in Preperimetric Glaucoma

Moreno-Montañés J (1), García-Nieva A (1), Osio IA (2), Guarnieri A (1), Morilla-Grasa A (2), García-Granero M (3), Antón A (2).

(1) Department of Ophthalmology, Clínica Universidad de Navarra, Medical School, University of Navarra, Pamplona, Spain.
(2) Institut Catalá de Retina, Universitat Internacional de Catalunya, Barcelona, Spain.
(3) Unidad de Estadística, Department of Biochemistry and Genetics, University of Navarra, Pamplona, Spain.

Revisão:Translational Vision Science & Technology

Data: 30/Nov/2018

Oftalmologia [ES]

PURPOSE:

To evaluate two glaucoma diagnostic calculators (GDC) in a group of eyes with preperimetric glaucoma (PPG).

METHODS:

All eyes (n = 265) included in this study had ocular hypertension with normal visual fields (VFs) on repeated VF tests. PPG was defined as progression in the Guided Progression Analysis software from Cirrus-optical coherence tomography (GPA-OCT). Three PPG types were defined according to the GPA-OCT software as follows: (1) GPA-OCT with one or more red boxes in two or more columns; (2) GPA-OCT with two or more red boxes in two or more columns; and (3) GPA-OCT with two or more red boxes in two or more columns (definition 2), and in the last scan one or more red box in the RNFL average or quadrants. Nonparametric tests, areas under the receiver operating characteristic curve (AUC), and Bland-Altman tests were assessed.

RESULTS:

Definitions one, two, and three were met by 44 (16.6%), 29 (10.9%), and 11 (4.2%) eyes, respectively. The GDC indices (means ± standard deviations) were, respectively, 14.49 ± 21.55% and 26.06 ± 22.50% using the combined and quantitative GDC (P < 0.001) in all eyes. Both GDC showed higher glaucoma probability in the PPG group (P < 0.04; combined GDC AUCs, 0.720-0.833; quantitative GDC AUCs, 0.700-0.839). GDC values were higher (P < 0.01) with greater GPA progression.

CONCLUSIONS:

The values of both GDC were higher in the PPG group than the ocular hypertension group. The GDC were higher when more columns in the GPA software indicated progression. Both GDC showed a similar ability to detect PPG.

TRANSLATIONAL RELEVANCE:

These calculators facilitate diagnosis of PPG in ocular hypertensive eyes.

CITAÇÃO DO ARTIGO  Transl Vis Sci Technol. 2018 Nov 30;7(6):13. doi: 10.1167/tvst.7.6.13

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra