Using NS5B Sequencing for Hepatitis C Virus Genotyping Reveals Discordances with Commercial Platforms

Chueca N (1), Rivadulla I (2), Lovatti R (3), Reina G (4), Blanco A (5), Fernandez-Caballero JA (1), Cardeñoso L (5), Rodriguez-Granjer J (6), Fernandez-Alonso M (4), Aguilera A (2), Alvarez M (1), Galán JC (3), García F (1).
(1) Servicio de Microbiología, Complejo Hospitalario Universitario Granada-Hospital San Cecilio, Instituto de Investigación Biosanitaria IBS, Granada, Spain.
(2) Departamento de Microbiología y Parasitología USC, Servicio de Microbiología Hospital Conxo-CHUS Santiago de Compostela, A Coruña, Spain.
(3) Servicio de Microbiología, Hospital Ramón y Cajal, Instituto Ramón y Cajal de Investigación Sanitaria (IRYCIS), Madrid, Spain; CIBER en Epidemiología y Salud Pública (CIBERESP), Madrid, Spain.
(4) Servicio de Microbiología, Clínica Universidad de Navarra, Navarra Institute for Health Research (IdiSNA), Pamplona, Spain.
(5) Servicio de Microbiología, Hospital la Princesa, Madrid, Spain.
(6) Servicio de Microbiología, Complejo Hospitalario Universitario Granada, Hospital Virgen de las Nieves, Granada, Spain. 

Revista: PloS One

Fecha: 01-abr-2016

Microbiología Clínica

RESUMEN

We aimed to evaluate the correct assignment of HCV genotypes by three commercial methods-Trugene HCV genotyping kit (Siemens), VERSANT HCV Genotype 2.0 assay (Siemens), and Real-Time HCV genotype II (Abbott)-compared to NS5B sequencing.

We studied 327 clinical samples that carried representative HCV genotypes of the most frequent geno/subtypes in Spain. After commercial genotyping, the sequencing of a 367 bp fragment in the NS5B gene was used to assign genotypes.

Major discrepancies were defined, e.g. differences in the assigned genotype by one of the three methods and NS5B sequencing, including misclassification of subtypes 1a and 1b. Minor discrepancies were considered when differences at subtype levels, other than 1a and 1b, were observed. The overall discordance with the reference method was 34% for Trugene and 15% for VERSANT HCV2.0.

The Abbott assay correctly identified all 1a and 1b subtypes, but did not subtype all the 2, 3, 4 and 5 (34%) genotypes. Major discordances were found in 16% of cases for Trugene HCV, and the majority were 1b- to 1a-related discordances; major discordances were found for VERSANT HCV 2.0 in 6% of cases, which were all but one 1b to 1a cases.

These results indicated that the Trugene assay especially, and to a lesser extent, Versant HCV 2.0, can fail to differentiate HCV subtypes 1a and 1b, and lead to critical errors in clinical practice for correctly using directly acting antiviral agents.

CITA DEL ARTÍCULO  PLoS One. 2016 Apr 20;11(4):e0153754. doi: 10.1371/journal.pone.0153754. eCollection 2016

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra