Study of the neutron field in the vicinity of an unshielded PET cyclotron

Méndez R, Iñiguez MP, Martí-Climent JM, Peñuelas I, Vega-Carrillo HR, Barquero R.
Dpto. Física Teórica, Atómica y Optica, Universidad de Valladolid, Spain

Revista: Physics in Medicine and Biology

Fecha: 07-nov-2005

Medicina Nuclear

The neutron field in the proximity of an unshielded PET cyclotron was investigated during 18F radioisotope production with an 18 MeV proton beam.

Thermoluminescent detector (TLD) models TLD600 and TLD700 as well as Bonner moderating spheres were irradiated at different positions inside the vault room where the cyclotron is located to determine the thermal neutron flux, neutron spectrum and dose equivalent. Furthermore, from a combination of measurements and Monte Carlo simulations the neutron source intensity at the target was estimated. The resulting intensity is in good agreement with the IAEA recommendations.

Neutron doses derived from the measured spectra were found to vary between 7 and 320 mSv per 1 microA h of proton-integrated current. Finally, gamma doses were determined from TLD700 readings and amounted to around 10% of the neutron doses.

CITA DEL ARTÍCULO  Phys Med Biol. 2005 Nov 7;50(21):5141-52

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra