Publicaciones científicas

Evaluation of CdZnTe as neutron detector around medical accelerators

Martín-Martín A, Iñiguez MP, Luke PN, Barquero R, Lorente A, Morchón J, Gallego E, Quincoces G, Martí-Climent JM.
Departamento de Física Teórica, Atómica y Optica, Universidad de Valladolid, Valladolid, Spain.

Revista: Radiation Protection Dosimetry

Fecha: 01-feb-2009

Medicina Nuclear

The operation of electron linear accelerators (LINACs) and cyclotrons can produce a mixed gamma-neutron field composed of energetic neutrons coming directly from the source and scattered lower energy neutrons. The thermal neutron detection properties of a non-moderated coplanar-grid CdZnTe (CZT) gamma-ray detector close to an 18 MV electron LINAC and an 18 MeV proton cyclotron producing the radioisotope (18)F for positron emission tomography are investigated.

The two accelerators are operated at conditions producing similar thermal neutron fluence rates of the order of 10(4) cm(-2) s(-1) at the measurement locations. The counting efficiency of the CZT detector using the prompt 558 keV photopeak following (113)Cd thermal neutron capture is evaluated and a good neutron detection performance is found at the two installations.

CITA DEL ARTÍCULO  Radiat Prot Dosimetry. 2009 Feb;133(4):193-9

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra