A new method to measure electron density and effective atomic number using dual-energy CT images

Garcia LI (1), Azorin JF, Almansa JF.
(1) Department of Oncology, Clínica Universidad de Navarra, University of Navarra, Av Pio XII s/n, Pamplona, Navarre, Spain 

Revista: Physics in Medicine and Biology

Fecha: 07-ene-2016

Radiofísica Oncología Radioterápica

The purpose of this work is to present a new method to extract the electron density ([Formula: see text]) and the effective atomic number (Z eff) from dual-energy CT images, based on a Karhunen-Loeve expansion (KLE) of the atomic cross section per electron.

This method was used to calibrate a Siemens Definition CT using the CIRS phantom. The predicted electron density and effective atomic number using 80 kVp and 140 kVp were compared with a calibration phantom and an independent set of samples.

The mean absolute deviations between the theoretical and calculated values for all the samples were 1.7 %  ±  0.1 % for [Formula: see text] and 4.1 %  ±  0.3 % for Z eff.

Finally, these results were compared with other stoichiometric method. The application of the KLE to represent the atomic cross section per electron is a promising method for calculating [Formula: see text] and Z eff using dual-energy CT images.

CITA DEL ARTÍCULO  Phys Med Biol. 2016 Jan 7;61(1):265-79. doi: 10.1088/0031-9155/61/1/265

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra