Juan Diego Azcona [SP] (a), Javier Burguete (b)
(a) Department of Oncology, Division of Radiation Physics, Clínica Universitaria, Universidad de Navarra, Avda. Pío XII, 36, 31008 Pamplona, Navarra, Spain
(b) Department of Physics and Applied Mathematics, Facultad de Ciencias, Universidad de Navarra, C/ Irunlarrea s/n, 31008 Pamplona, Navarra, Spain

Magazine: Medical Physics

Date: Dec 21, 2007

Radiophysics and Radiological Protection [SP] Radiation Oncology

The number of intensity modulated radiation therapy (IMRT) procedures is continuously growing worldwide and it is necessary to develop tools for patient specific quality assurance (QA) that avoid using machine time that could be employed in treating additional patients.

One way of achieving this goal is to perform a multileaf collimator quality assurance periodically in the linear accelerator and check the treatment planning system (TPS) calculation by employing an independent calculation system. Within the work frame of the pencil beam kernel approach, a new system was developed for obtaining an experimental kernel. This new technique is based on a deconvolution procedure using the Hankel transform.

The resulting kernel is obtained in a way completely independent of those employed in commercial treatment planning systems, usually calculated by Monte Carlo simulations. Also provided are comparisons between calculated and measured doses with radiographic film, linear array of diodes, and ionization chamber.

Measurements taken in polystyrene and water for clinical IMRT plans demonstrate that this method can calculate IMRT dose distributions, as well as treatment times, with great accuracy. Apart from other applications, it can be used as a double-check algorithm for IMRT QA.

CITATION  Med Phys. 2008 Jan;35(1):248-59

you mayBE INTERESTED

WHAT TECHNOLOGY
DO WE USE?

The Clínica is the greater private hospital with technological equipment of Spain, all in a single center.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OUR
PROFESSIONALS

The professionals of the Clínica perform continuous research and training, always to the benefit of the patient.

Imagen profesionales de la Clínica Universidad de Navarra

WHY CHOOSE
THE CLINICA?

Learn why we are different from other healthcare centers. Quality, speed, comfort and results.

Imagen del edificio de la Clínica Universidad de Navarra