Publicações científicas

Sampling size in the verification of manufactured-supplied air kerma strengths

Ramos L.I. [ES], Martínez Monge R.
Department of Oncology, Clínica Universitaria de Navarra, University of Navarra, Avda Pío XII s/n, Pamplona, Navarra, Spain.

Revisão:Medical Physics Online

Data: 1/Nov/2005

Oncologia Radioterapêutica

Quality control mandate that the air kerma strengths (S(K)) of permanent seeds be verified, this is usually done by statistics inferred from 10% of the seeds. The goal of this paper is to proposed a new sampling method in which the number of seeds to be measured will be set beforehand according to an a priori statistical level of uncertainty.

The results are based on the assumption that the S(K) has a normal distribution. To demonstrate this, the S(K) of each of the seeds measured was corrected to ensure that the average S(K) of its sample remained the same. In this process 2030 results were collected and analyzed using a normal plot. In our opinion, the number of seeds sampled should be determined beforehand according to an a priori level of statistical uncertainty.

CITAÇÃO DO ARTIGO  Med Phys. 2005 Nov;32(11):3375-8

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra