Radiation dose produced by patients during radiopharmaceutical incorporation in nuclear medicine diagnostic procedures

Morán V (1), Prieto E [ES] (2), García-García B (3), Barbés B [ES] (2), Ribelles MJ (3), Richter JÁ [ES] (4), Martí-Climent JM [ES] (5).
(1) Servicio de Radiofísica y Protección Radiológica, Clínica Universidad de Navarra, Pamplona, España.
(2) Servicio de Radiofísica y Protección Radiológica, Clínica Universidad de Navarra, Pamplona, España; Instituto de Investigación Sanitaria de Navarra (IDISNA), Pamplona, España.
(3) Servicio de Medicina Nuclear, Clínica Universidad de Navarra, Pamplona, España.
(4) Instituto de Investigación Sanitaria de Navarra (IDISNA), Pamplona, España; Servicio de Medicina Nuclear, Clínica Universidad de Navarra, Pamplona, España.
(5) Servicio de Radiofísica y Protección Radiológica, Clínica Universidad de Navarra, Pamplona, España; Instituto de Investigación Sanitaria de Navarra (IDISNA), Pamplona, España 

Revisão:Revista Española de Medicina Nuclear e Imagen Molecular

Data: 29/Dez/2015

Medicina Nuclear [ES] Radiofísica [ES]

OBJECTIVES

The aim of this study was to assess the dose received by members of the public due to close contact with patients undergoing nuclear medicine procedures during radiopharmaceutical incorporation, and comparing it with the emitted radiation dose when the test was complete, in order to establish recommendations.

MATERIAL AND METHODS

A prospective study was conducted on 194 patients. H*(10) dose rates were measured at 0.1, 0.5, and 1.0m after the radiopharmaceutical administration, before the image acquisition, and at the end of the nuclear medicine procedure. Effective dose for different close contact scenarios were calculated, according to 95th percentile value (bone scans) and the maximum value (remaining tests).

RESULTS

During the radiopharmaceutical incorporation, a person who stays with another injected patient in the same waiting room may receive up to 0.59mSv. If the patient had a medical appointment, or went to a restaurant or a coffee shop, members of the public could receive 23, 43, and 22μSv, respectively. After finishing the procedure, these doses are reduced by a factor 3. In most of the studies, the use of private instead of public transport may reduce the dose by more than a factor 6.

CONCLUSION

It is recommended to increase the distance between the patients during the radiopharmaceutical incorporation and to distribute them according to the diagnostic procedure. Patients should be encouraged to use private instead of public transport. Depending on the number of nuclear medicine outpatients per year attended by a physician, it could be necessary to apply restrictions.

CITAÇÃO DO ARTIGO  Rev Esp Med Nucl Imagen Mol. 2015 Dec 29. pii: S2253-654X(15)00145-6. doi: 10.1016/j.remn.2015.10.004.

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra