Objective quantification of posterior capsule opacification after cataract surgery, with optical coherence tomography.

Moreno-Montañés J [ES], Alvarez A, Maldonado MJ.
Department of Ophthalmology, Clínica Universitaria, Facultad de Medicina, Universidad de Navarra, Pamplona, Spain.

Revisão:Investigative Ophthalmology & Visual Science

Data: 1/Nov/2005

Oftalmologia [ES]

PURPOSE
To evaluate posterior capsule opacification (PCO) in humans after cataract surgery with intraocular lens (IOL) implantation, by using optical coherence tomography (OCT-1).

METHODS
Sixty-six eyes with PCO and 20 eyes with a normal posterior capsule were analyzed. A 3-mm-long horizontal scan of the posterior capsule was obtained. Measurements at three points and their average were recorded. Intraoperator and interoperator reliabilities were assessed. Investigated was peak intensity (PI) and posterior capsule thickening (PCT), with PCT indicating the distance between two reflectivity spikes, with an approximate axial resolution of 10 microm. Results were compared with visual acuity (VA) and PCO type.

RESULTS
Intraoperator reliability was 0.59 and 0.97 for average PI and PCT, respectively. The interoperator concordance correlation coefficient was 0.70 and 0.82 for average PI and PCT, respectively. Median (interquartile range) intensities of the reflectivity spike were 16.88 (dB) (range, 12.88-20.41) and 11.9 (8.58-14.28), respectively, in the PCO and control eyes (P = 0.001). PCT was found in PCO eyes (median: 86.13 microm; range, 46.33-115.33), whereas no second spike appeared in control eyes (P = 0.001). The area under the receiver operating characteristic curve of the average PCT for differentiating pearl-type from fibrosis-type PCO was 0.87 (P = 0.001). For a cutoff point of 55.3 microm, the sensitivity was 97.5%, and the specificity was 69%. Worse VA correlated significantly only with larger PCT (r(o) = 0.66; P = 0.01).

CONCLUSIONS
OCT-1 appears useful to quantitate PCO. In addition, this system seems to discriminate between different types of PCO. PCT may be a previously unrecognized factor in VA degradation.

CITAÇÃO DO ARTIGO  Invest Ophthalmol Vis Sci. 2005 Nov;46(11):3999-4006

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra