RESUMEN
Computerized electronystagmography (ENG) is the analysis of the eye movement recordings using a computer program to identify nystagmus and quantitate response parameters. The results of 100 consecutive bithermal caloric tests of patients complaining of peripheral vertigo were analyzed by computer.

All computer analyses were reviewed by authors and either were deemed accurate as made by the computer or were corrected manually. Final test interpretations with and without manual intervention were compared. In 48 tests the computer analysis was not modified. In 20 tests canal paresis and directional preponderance changed after manual review, but not the final diagnosis. In the remaining 20 tests, manual review yielded a different final diagnosis.

Computerized ENG provides rapid and accurate vestibular studies, but recordings should be reviewed because system algorithms sometimes identify other ocular movements as nystagmus.

CITAÇÃO DO ARTIGO Acta Otorrinolaringol Esp. 1996 Sep-Oct;47(5):354-8

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra