An independent algorithm to check the monitor units calculation in radiosurgery

Luis Isaac Ramos [ES], Rafael Martínez Monge, José Javier Aristu [ES], and Leire Arbea Moreno [ES]
Department of Oncology, Clínica Universitaria de Navarra, University of Navarra,Avda Pío XII s/n. Pamplona, Navarre, Spain, Pc 31008

Revisão:Medical Physics

Data: 11/Dez/2007

Oncologia Radioterapêutica

Double checking of the monitor units (MU) is an important step in the quality assurance (QA) process in radiosurgery. In this paper we propose the use of an independent algorithm constructed using the ellipsoid which best fits the measurements taken with the bubble head frame.

The monitor units calculated by this independent algorithm and the commercial planning system were compared in 40 patients treated with radiosurgery (57 isocenters, 320 arcs). The average relative difference was -0.2% +/- 2.1 (k=1). These results are better for the variance, -0.4% +/- 1.8 (k=1), when all the depths of the bubble head frame are measured and no arcs are calculated by extrapolation or when only one of these factors appear.

If there are missing values in the bubble head frame measurements and the model is extrapolated, the variance of the results is greater, 0.4% +/- 3.9 (k=1). The algorithm is reliable as a QA tool for linac radiosurgery.

CITAÇÃO DO ARTIGO  Med Phys. 2008 Jan;35(1):48-51

talvezlhe interesse

QUE TECNOLOGIA UTILIZAMOS? 

A Clínica é o hospital privado com maiores recursos tecnológicos de Espanha, tudo num único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OS NOSSOS
PROFISSIONAIS

Os profissionais da Clínica realizam um trabalho contínuo de investigação e formação, sempre em benefício do paciente.

Imagen profesionales de la Clínica Universidad de Navarra

RAZÕES PARA VIR
À CLÍNICA

Conheça porque é que somos diferentes em relação a outros centros sanitários. Qualidade, rapidez, comodidade e resultados.

Imagen del edificio de la Clínica Universidad de Navarra