Sampling size in the verification of manufactured-supplied air kerma strengths

Ramos L.I., Martínez Monge R.
Department of Oncology, Clínica Universitaria de Navarra, University of Navarra, Avda Pío XII s/n, Pamplona, Navarra, Spain.

Revista: Medical Physics Online

Fecha: 01-nov-2005

Oncología Radioterápica

Quality control mandate that the air kerma strengths (S(K)) of permanent seeds be verified, this is usually done by statistics inferred from 10% of the seeds. The goal of this paper is to proposed a new sampling method in which the number of seeds to be measured will be set beforehand according to an a priori statistical level of uncertainty.

The results are based on the assumption that the S(K) has a normal distribution. To demonstrate this, the S(K) of each of the seeds measured was corrected to ensure that the average S(K) of its sample remained the same. In this process 2030 results were collected and analyzed using a normal plot. In our opinion, the number of seeds sampled should be determined beforehand according to an a priori level of statistical uncertainty.

CITA DEL ARTÍCULO  Med Phys. 2005 Nov;32(11):3375-8

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra