Rebound tonometer compared with goldmann tonometer in normal and pathologic corneas

Moreno-Montañés J, García N, Fernández-Hortelano A, García-Layana A.
Department of Ophthalmology, Clínica Universitaria, Universidad de Navarra, Pamplona, Spain.

Revista: Cornea

Fecha: 01-may-2007

Oftalmología

PURPOSE
To compare the intraocular pressure (IOP) measurements obtained with the rebound tonometer (RT) and the Goldmann handheld tonometer (GT) in normal and altered corneas.

METHODS
A total of 208 normal corneas and 50 corneas with pathologies were included in this prospective study. All measurements were randomly obtained by 1 observer. The medians and interquartile range (IR) for both tonometers were compared. The median differences were assessed in IOP groups. Agreement between the tonometers was calculated using the Bland-Altman method.

RESULTS
The median IOP in all eyes was 17 mm Hg (IR, 13-22 mm Hg) with the RT and 16 mm Hg (IR, 13-21 mm Hg) with the GT (P < 0.001). The correlation was excellent between tonometers (r2 = 0.86; P < 0.001). The minimal differences between the two were obtained from 10 to 20 mm Hg (GT). The Bland-Altman scatterplot obtained good agreement between the instruments. In normal corneas, the median difference was < or =2 mm Hg in 77.4% of cases. In the altered corneas, the median difference was < or =2 mm Hg in 73% of cases (P = 0.21 compared with the normal group). In 10% and 2% of cases, the IOP could not be measured using the GT and RT, respectively.

CONCLUSIONS
The results were similar for both tonometers. In the altered corneas, the IOP could be difficult to obtain with the GT because of distorted half-circles. The 1-mm-diameter disposable RT tip facilitated obtaining measurements without anesthetic drops, which avoids infections. The RT could be useful in routine clinical settings when measuring IOP in corneas with pathologies.

CITA DEL ARTÍCULO  Cornea. 2007 May;26(4):427-30

tal vezLE INTERESE

¿QUÉ TECNOLOGÍA UTILIZAMOS?

La Clínica es el hospital privado con mayor dotación tecnológica de España, todo en un único centro.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

NUESTROS
PROFESIONALES

Los profesionales de la Clínica realizan una labor continuada de investigación y formación, siempre en beneficio del paciente.

Imagen profesionales de la Clínica Universidad de Navarra

POR QUÉ VENIR
A LA CLÍNICA

Conozca por qué somos diferentes a otros centros sanitarios. Calidad, rapidez, comodidad y resultados.

Imagen del edificio de la Clínica Universidad de Navarra