Electric parameters optimization in spinal cord stimulation. Study in conventional nonrechargeable systems

Abejón D, Cameron T, Feler C.
Head of Unit Pain. Hospital Universitario Puerta de Hierro Majadahonda, Madrid, Spain; Biomedical Engineering Program, VP Clinical Research, Neuromodulation Division, St. Jude Medical, TX, USA; University of Tennessee Center for Health Sciences, Semmes-Murphey Neurologic & Spine Institute, Memphis, TN, USA; and Head of Unit Pain. Clínica Universitaria de Navarra, Pamplona, Spain.

Magazine: Neuromodulation

Date: Oct 1, 2010

Pain Unit [SP]

Background
Spinal cord stimulation devices provide a means of creating an electric field. The parameters used to produce this electric field are: pulse amplitude, pulse width (Pw), and pulse frequency (F).

Aims
The purpose is to document the effects that the various stimulus parameters have on patient perception of paresthesia and the relationship that this perception has on pain relief.

Methods
Stimulus parameters were varied independently keeping the electrode polarity constant while recording stimulation thresholds. The Pw was varied from 195 to 300 µsec while maintaining the frequency at 50 Hz. The F was varied from 10 to 100 Hz while maintaining the Pw at 300µsec.

We also measured the paresthesia coverage percentage and the subjective perception of quality reported by the patients with each one of the parameter changes. Results:  There was a statistically significant correlation between Pw and all the stimulation thresholds. As for the therapeutic range, the differences observed also were statistically significant. Pw variation did not produce significant differences in coverage and subjective quality of the paresthesia. The perception threshold did not vary significantly with F changes. However, F significantly affected both coverage of the painful area and paresthesia perception quality.

Conclusions
In the usual Pw ranges, it seems that its usefulness is limited to obtaining finer adjustments in the stimulation amplitude. Frequency management may be significantly useful to get a wider coverage of the stimulated area.

CITATION  Neuromodulation. 2010 Oct;13(4):281-7. doi: 10.1111/j.1525-1403.2010.00290.x

you mayBE INTERESTED

WHAT TECHNOLOGY
DO WE USE?

The Clínica is the greater private hospital with technological equipment of Spain, all in a single center.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OUR
PROFESSIONALS

The professionals of the Clínica perform continuous research and training, always to the benefit of the patient.

Imagen profesionales de la Clínica Universidad de Navarra

WHY CHOOSE
THE CLINICA?

Learn why we are different from other healthcare centers. Quality, speed, comfort and results.

Imagen del edificio de la Clínica Universidad de Navarra