An independent algorithm to check the monitor units calculation in radiosurgery

Luis Isaac Ramos [SP], Rafael Martínez Monge, José Javier Aristu [SP], and Leire Arbea Moreno [SP]
Department of Oncology, Clínica Universitaria de Navarra, University of Navarra,Avda Pío XII s/n. Pamplona, Navarre, Spain, Pc 31008

Magazine: Medical Physics

Date: Dec 11, 2007

Radiation Oncology

Double checking of the monitor units (MU) is an important step in the quality assurance (QA) process in radiosurgery. In this paper we propose the use of an independent algorithm constructed using the ellipsoid which best fits the measurements taken with the bubble head frame.

The monitor units calculated by this independent algorithm and the commercial planning system were compared in 40 patients treated with radiosurgery (57 isocenters, 320 arcs). The average relative difference was -0.2% +/- 2.1 (k=1). These results are better for the variance, -0.4% +/- 1.8 (k=1), when all the depths of the bubble head frame are measured and no arcs are calculated by extrapolation or when only one of these factors appear.

If there are missing values in the bubble head frame measurements and the model is extrapolated, the variance of the results is greater, 0.4% +/- 3.9 (k=1). The algorithm is reliable as a QA tool for linac radiosurgery.

CITATION  Med Phys. 2008 Jan;35(1):48-51

you mayBE INTERESTED

WHAT TECHNOLOGY
DO WE USE?

The Clínica is the greater private hospital with technological equipment of Spain, all in a single center.

Imagen de un PET, tecnología de vanguardia en la Clínica Universidad de Navarra

OUR
PROFESSIONALS

The professionals of the Clínica perform continuous research and training, always to the benefit of the patient.

Imagen profesionales de la Clínica Universidad de Navarra

WHY CHOOSE
THE CLINICA?

Learn why we are different from other healthcare centers. Quality, speed, comfort and results.

Imagen del edificio de la Clínica Universidad de Navarra